Trying Out the Instrument
R&S
®
RTO
69Getting Started 1316.0833.02 ─ 10
8. Select the file type: "*.bin".
9. Tap "Save".
The waveform data is saved to the files Waveform1.Wfm.bin and
Waveform1.bin in the selected directory.
Saving data of an acquisition series
1. Press the FILE key on the front panel.
2. Select the "Waveforms /Results" tab.
3. Set the export scope of the waveform:
a) Check the "Source" and set "Scope" to "Full Waveform".
b) Tap "Data logging" to enable export all waveforms of a running acquisition.
c) Enter "Acq count" = 10, the number of subsequent waveforms to be saved.
4. Tap "Start Export" to save the waveforms to the file that is named under "Save to
file".
Saving measurement results
1. Perform a measurement as described in chapter 4.7.2, "Performing an Amplitude
Measurement", on page 52.
2.
Tap the icon in the result box.
3. Tap the "Result export" button in the "Setup" tab.
4. On the "Numeric Results" tab, select the results to be saved.
Saving Data