11
ELECTROMAGNETIC COMPATIBILITY
This equipment has been tested and found to comply with the limits for medical devices to the
IEC60601-1-2 Electromagnetic Compatibility standard. These limits are designed to provide
reasonable protection against harmful interference in a typical medical installation. This equipment
generates, uses, and can radiate radio frequency energy and, if not installed according with the
instructions, may cause harmful interference to other devices in the vicinity. However, there is no
guarantee that interference will not occur in a particular installation. If this equipment does cause
harmful interference to other devices, which can be determined by turning the equipment off and on,
the user is encouraged to try to correct the interference by one or more of the following measures:
Reorient or relocate the receiving device.
Increase the separation distance between the equipment.
Connect the equipment into an outlet on a circuit different from that which the other device(s)
are connected.
Consult with SeQual Technical Support for help.
Guidance and Manufacturer's declaration - electromagnetic
emissions
The SeQual Technologies Eclipse Oxygen System is intended for use in the electromagnetic
environment specified below. The customer or the user of the SeQual Technologies Eclipse
Oxygen System should assure that it is used in such an environment.
Emissions Test
Compliance Electromagnetic environment - guidance
RF emissions
EN 55011
Group 1
The SeQual Technologies Eclipse Oxygen System
uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby
electronic equipment.
RF emissions
EN 55011
Class B
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
The SeQual Technologies Eclipse Oxygen System
is suitable for use in all establishments, including
domestic establishments and those directly
connected to the public low - voltage power supply
network that supplies buildings used for domestic
purposes.