Introduction
Maintaining the Eclipse will ensure longer life and higher performance. Periodic maintenance is
recommended.
Periodic Maintenance List
Perform the following maintenance procedures, as needed. The frequency of the periodic maintenance
should be based on the environment in which the Eclipse is used within.
Check and clean air inlet filter. Replace as needed
Check Power Cartridge
Inspect AC and DC Power Supply plugs and cords for damage. Replace as needed
Verify that the AC and DC Power Supplies operate with the Eclipse
Readjust the Pulse Flow Sensitivity Adjustment, in between patients as needed
Check 9V battery. Replace if needed
Check cart for functioning wheels and telescoping handle. Replace as needed
Read and record hour meter
Check flow rate, purity, Pulse Mode and alarm functions
Check compressor intake filter, replace as needed.
Perform electrical safety test (Required only for PN 5701, Hospital Grade Power Cord)
Replace HEPA filter
Checking and Replacing the Power Cartridge
Typically, the Power Cartridge will require no routine maintenance beyond cleaning by the patient.
SeQual recommends that the Power Cartridge be replaced if there is physical damage to the
cartridge enclosure or connector, or if the excursion time is significantly less than published values at
a given flow rate.
Proper Care for the Power Cartridge
Avoid high temperatures
Avoid cold temperatures
Do not drop Power Cartridge
No not poke objects into contacts
NOTE:
Dispose the Power Cartridge in accordance with local disposal
regulations or return to SeQual.