8
Eclipse Oxygen System Specifications
Oxygen Concentrator
Dimensions (H x W X D)
19.3 x 12.3 x 7.1 inches
(49.0cm x 31.2cm x 18.0cm)
Weight
With Power Cartridge
Without Power Cartridge
17.9 pounds (8.1 kg)
14.5 pounds (6.6 kg)
Flow Settings
Continuous Flow
Pulse Flow
0.5 to 3.0 LPM, 0.5 LPM increments
Settings of 1.0 to 6.0, in 0.5 increments
Continuous Flow Accuracy +/- 10% or 200ml/min, whichever is greater
Oxygen concentration 90% +/- 3% for all flow settings
Maximum System Pressure
15 psig (103.5 kPa)
Oxygen Output Pressure 5.0 psig (34.5 kPa) nominal
Oxygen Concentration
Indicator
Green Light = Normal Operation
Yellow Light = Warning or Caution, less than 85% ± 3%
Red Light Flashing = Abnormal Operation, less than 70% ± 5%
Nominal Sound Level
3 LPM Continuous Flow
0.5LPM Continuous Flow
48 dB(A)
40 dB(A)
Operating Environment
Temperature
Humidity
50º F to 104º F (+10° to 40°C)
10% to 95%, Non-condensing, 82.4°F (28°C) Maximum Dew
point
Storage Environment
Temperature
Humidity
-4º F to 140º F (-20° to 60°C)
Up to 95% Non-condensing
Nominal Power
0.5 LPM Continuous flow
3 LPM Continuous Flow
1.0 Pulse Flow Setting
6.0 Pulse Flow Setting
48 Watts
145 Watts
45 Watts
95 Watts
Nominal Power Cartridge
operating time
Continuous Flow
1.0 LPM
3.0 LPM
Pulse Flow
1.0
6.0
4.4 hours
1.3 hours
4.4 hours
2.1 hours
Continuous Flow Indication
Expressed in liters per minute (LPM)
Note: Times will decrease with
higher bolus size, breath rate,
ambient temperature, power
cartridge age and use over time.