51
Remove and Replace HEPA Filter
Replace the HEPA filter as needed. To replace the HEPA filter, follow these steps.
WARNING:
DO NOT
use any petroleum based or other lubricants. A
spontaneous and violent ignition may occur if oil, grease or other
petroleum substances come into contact with oxygen under
pressure. Keep these substances away from the oxygen system,
tubing and connections and any other oxygen source.
NOTE:
Always cut head of cable ties to avoid damaging the tubing.
1. Cut the three cable ties on the silicone tubes attached to the HEPA filter.
Figure 20: Removing the HEPA Filter.
2. Disconnect both tubes from the filter nipples.