System Troubleshooting and Alarms
Troubleshooting
SYMPTOM PROBABLE CAUSE REMEDY
No Power Cartridge installed Install charged Power Cartridge or
plug into external power
Power Cartridge is discharged Plug into external power
Eclipse does not power on
when On/OFF button is
pressed
Malfunction Contact SeQual
Blocked or kinked tubing Check tubing or replace
Filter Blocked Clean Air Inlet Filter
Humidifier Check humidifier attachment and
tubing
Inlet or Exhaust Vent obstructed Ensure proper ventilation
Eclipse not ON Turn Eclipse ON
Tubing or cannula is not properly
connected or kinked
Check tubing, cannula and
connections
Power Cartridge is discharged or
no external power is present
Install charged Power Cartridge or
attach external power
No Oxygen
Malfunction Contact SeQual
Restriction in tubing or humidifier Repair or replace tubing or
humidifier
Low Oxygen Concentration
Filter restricted Clean or replace air inlet filter
Malfunction Contact SeQual
Restriction in humidifier or tubing Repair or replace humidifier or
tubing
Low Oxygen Flow
Filter Blocked Clean or replace air inlet filter
Malfunction Contact SeQual
Tubing/cannula longer than 7
feet (2.1m)
Attach 7 foot (2.1m) tubing/cannula
Humidifier attached Remove humidifier
No inspiration detected Increase Pulse Flow Sensitivity by
changing sensitivity value to a lower
number
No Oxygen delivered in Pulse
Mode
No inspiration detected Contact SeQual
Auto-triggers in Pulse Mode
(random pulsing)
Pulses without patient
inhalations
Decrease Pulse Flow Sensitivity by
changing sensitivity value to a higher
number
Power Cartridge Status
Gauge never indicates fully
charged
Power Cartridge needs replaced Replace with new Power Cartridge