WARNING:
When using the Eclipse in a vehicle, always check the cannula to
make sure it is not pinched or occluded.
WARNING:
• Avoid placing the device in direct sunlight.
• Do not store the Eclipse in a vehicle where the device may be
subject to extreme temperatures. Extreme heat or cold may impair
operation and damage the device and degrade the Power
Cartridge
.
Use the following table to determine whether your patient’s vehicle DC power is sufficient to supply the
necessary current to sustain your patient’s desired flow rate.
It is recommended to use Pulse Flow if it is prescribed by the clinician.
The DC-DC Power Supply compatibility is as follows:
DC Power Supply
Operation
Flow Mode
Flow
Setting
15 Amp Mode 10 Amp
Mode
Continuous 0.5 LPM Yes Yes
Continuous 1.0 LPM Yes Yes
Continuous 1.5 LPM Yes Yes
Continuous 2.0 LPM Yes Yes
Continuous 2.5 LPM Yes
No
Continuous 3.0 LPM Yes
No
Pulse 1 Yes Yes
Pulse 2 Yes Yes
Pulse 3 Yes Yes
Pulse 4 Yes Yes
Pulse 5 Yes Yes
Pulse 6 Yes Yes
Table 2: Flow Modes for DC Power Supply Operation.
Traveling by Air
When traveling by air, instruct your patient to remove the Power Cartridge from the Eclipse before
checking the unit as checked baggage or cargo. Transport of the Eclipse as cargo or checked baggage
with the Power Cartridge installed is prohibited by international air cargo regulations.