64
Condition
Green
Indicator
Yellow
Indicator
Red
Indicator
Audible
Alarm
Alarm
Code
What to do
“Power Cartridge
Shutdown or hot
Power Cartridge ”
(with no external
power present)
Off Off Blinking On
Instruct patient to seek an
external power source. Remove
Power Cartridge and allow to cool
to room temperature. If Power
Cartridge malfunction persists,
service is required.
“Loss of External
Power” (without
charged Power
Cartridge
installed)
*Note: Stars will
be flashing in the
LCD ********”
Off Off Blinking On 40
Instruct patient seek an
alternative oxygen supply until
external power is restored or
install a charged Power Cartridge.
SHUTDOWN CONDITIONS
“System Fault” Off Off On On
Instruct patient to use back up
oxygen supply. Service is
required.