Feature 1 Separation Certification / Area Classification
1.6
04 (1100H) 100m maximum Transducer unit CENELEC Zone 1, Control unit Safe Area
only
06 500m maximum Transducer CENELEC Zone 1, Interface and Control unit
Safe Area only
16 500m maximum Transducer CSA Div1, Interface and Control unit CSA Div 2
26 500m maximum Transducer FM Div1, Interface and Control unit FM Div 2
06 (1100H) 500m maximum Transducer unit CENELEC Zone 1, Interface and Control unit.
Safe Area only
The 1100H is available in configuration codes 04 and 06 only.
With the exception of the close coupled units (codes 01 and 21) the configurations described
require interconnecting cables between the various units. This is not supplied unless
specifically requested since individual installation requirements vary. Cable specifications and
requirements are detailed in Sections 2.8 and 2.9 of this manual.
For codes 02 and 22 a terminal box is fitted to the Transducer unit allowing the Control unit to
be separated by a cable length of 100 metres maximum.
For codes 03 and 23 an Interface unit is close coupled to the Transducer unit allowing the
Control unit to be separated by a cable length of 500 metres maximum.
For codes 04, 14 and 24, two terminal boxes are fitted to the Transducer unit allowing the
Control unit to be separated by a cable length of 100 metres maximum.
For codes 06, 16 and 26, two terminal boxes are fitted to the Transducer unit allowing an
Interface unit to be separated by a cable length of 100 metres maximum and the Control unit to
be separated by a further cable length of 500 metres maximum.