3.16
TABLE 3.1 CALIBRATION FAILURE DISPLAYS
DISPLAY DESCRIPTION
0.01 77
0.02 77
10.01 77*
100.01 77*
0.06 77
0.07 77
0.77 77
0.01 78
0.02 78
0.04 78
0.06 78
0.07 78
0.78 78
Zero calibration successful
Zero calibration unsuccessful due to the zero gas being
outside the limits.
Both Zero and Span successfully calibrated.
Refer to 0.02 78
Unable to perform a manual calibration because analyser
is in the auto-calibration mode.
Unable to perform a manual calibration because analyser
is in the first stage of pressure compensation calibration.
No previous zero calibration done.
Span calibration successful.
Span calibration unsuccessful because:
1. Out of limits
2. Span gas set point less than or equal to 1.0%
more positive than zero gas set point.
3. Span reading less than or equal to 1.0% more
positive than zero gas reading.
4. Span calibration factor out of range.
Span calibration aborted because no valid previous zero
calibration.
Unable to perform a manual calibration because analyser
is in auto-calibration mode.
Unable to perform a manual calibration because analyser
is in the first stage of pressure compensation calibration.
No previous span calibration done.
* 10.01 77 and 100.01 77 are not available on software versions 0110067X and 0110068X.
Note: Span calibration will not be allowed to be performed if the previous zero calibration
was unsuccessful.