Plasma Gas Analyzer
9 02001001A_9
2 SPECIFICATION
WARNING
The protection, accuracy, operation and condition of the equipment
may be impaired if the analyzer is not installed in accordance with the
requirements of this and subsequent sections of the manual.
WARNING
Not suitable for flammable samples.
Never pressurise the analyzer.
Always remove vent plug before connecting the sample inlet.
(Height x Width x Depth):
133 mm x 483 mm x 457 mm
(5.25 ins x 19 ins x 18 ins)
15 kg
2.2 Environmental limits
Equipment is suitable for indoor use only
Ambient temperature range
Operating ambient pressure range:
101.3 kPa ± 10% (1.013 bar ± 10%)
Operating ambient humidity range:
0 to 95% RH, non-condensing
Operating altitude range: