Servomex SERVOPRO Plasma - Inspecting;Replacing the Fuse (When Necessary)

Servomex SERVOPRO Plasma
82 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Plasma Gas Analyzer
52 02001001A_9
6.2 Inspecting/replacing the fuse (when necessary)
WARNING
Disconnect power cord from the analyzer before replacing the fuse.
If you do not, there will be a danger of injury or death from electric
shock.
WARNING
Fire Hazard, only use the same type and
rated fuse as recommended.
If you think that an electrical supply fuse has failed, use the following procedure to
inspect the fuses and replace it if necessary:
1. Remove power cord.
2. Pry door open at socket with a suitable flat head screwdriver.
3. Lift and swing door into socket.
4. Lift fuse holder out of housing.
5. Install one (1) AG fuse or two (2) metric fuses*.
6. Replace fuse holder into housing.
7. Swing and snap door back in place.
*Install fuses on one side only, do not install both AG and metric fuses at the
same time.