Plasma Gas Analyzer
16 02001001A_9
4.2 Electrical Installation
4.2.1 Electrical Safety
WARNING
Ensure that the electrical installation of the analyzer conforms with all
applicable local and national electrical safety requirements.
WARNING
Obey the safety instructions given below when you install the analyzer;
if you do not, the analyzer warranty may be invalidated, the analyzer
may not operate correctly, or it may be damaged.
The electrical supply coupler or plug must be easily accessible for
disconnection from the electrical supply.
Ensure that your electrical supply can provide the necessary maximum
power consumption: refer to Section 10
All signal and electrical supply cables must be rated for temperatures
of 70 °C or higher.
Ensure that the cables that you connect to the analyzer are routed so that
they do not present a trip hazard.
When you carry out insulation testing, disconnect all cables from the
analyzer.
Ensure the analyzer is provided with a sound earth connection via the
electrical supply plug (Figure 2 Item 3).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Servomex SERVOPRO Plasma and is the answer not in the manual?

Servomex SERVOPRO Plasma Specifications

General IconGeneral
BrandServomex
ModelSERVOPRO Plasma
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals