Plasma Gas Analyzer
25 02001001A_9
4.3 Connect the sample/ auto calibration (optional) gas pipeline(s)
WARNING
Sample and calibration gases may be toxic or asphyxiant:
Ensure that the external connections are leak free at full
operating pressure before you use sample or calibration
gases.
Ensure that the sample outlet pipes are vented to an area
where the gases will not be a hazard to people.
Ensure that the analyzer is used in a sufficiently
well-ventilated environment, to prevent the build-up of toxic
gases.
Ensure that the pipes that you connect to the analyzer are routed
so that they do not present a hazard to people.
WARNING
The analyzer is not suitable for use with corrosive samples.
WARNING
Ensure the connectors used to connect to the analyzer are
compatible with the connectors on the analyzer.
Connect your sample/calibration gas inlet and outlet pipelines to the inlets and outlets on
the rear of the analyzer (see Figure 2 - REAR OF THE ANALY). The sizes of the fittings
are 1/8” compression fittings:
Refer to Section 2.4 and 2.5 for sample gas and calibration gas requirements. Locate
your gas selection valves as close as possible to the analyzer.
CAUTION
The Vent connection of the analyzer must be at atmospheric pressure at all
times. Pressurising the vent side of the analyzer could cause permanent damage
to the quartz cell and flow module.