4 INSTALLATION AND SET-UP
WARNING
The analyzer must be installed by a suitably skilled and
competent person. The following procedure must be followed
to prevent a hazard.
WARNING
The analyzer is only suitable for installation in unclassified areas.
CAUTION
Do not install the analyzer in a frame/rack or on a bench which is subject to
high levels of vibration or sudden jolts. If you do, sample measurements
may not be accurate, or the analyzer may be damaged.
CAUTION
Do not install analyzer in an area where a strong electromagnetic field is
present. Never use radio transmitters near the analyzer.
CAUTION
To avoid damage to analyzer, always leave the protection caps on gas
connections until clean gas is ready to flow in the analyzer.
CAUTION
The vent connection of the analyzer must be at atmospheric pressure all the
time. Pressurizing the vent side of analyzer could cause the quartz cell to
crack and also damage the flow module.
CAUTION
Do not expose the analyzer to high level of direct sunlight.