Servomex SERVOPRO Plasma - APPENDIX 4: INTERFACE FOR EXTERNAL VALVE BOX (VB); 4 INTERFACE FOR EXTERNAL VALVE BOX (VB)

Servomex SERVOPRO Plasma
82 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Plasma Gas Analyzer
73 02001001A_9
CAUTION
The metal based trap is made with ¼” O.D. S.S. tubing. Don’t allow air or
high level of O2 to flow in it because excessive heat would be generated.
Install the trap only when all lines are purged. This trap must be installed to
get good results when operating under helium. This trap can be shipped
back to Servomex for proper regeneration. We strongly recommend
keeping one in stock.