52 GMS800 · Operating Instructions · 8013025 V1.2 · © SICK MAIHAK GmbH
Maintenance
Subject to change without notice
8.1 Maintenance plan
8.1.1 Maintenance by the user
8.1.2 Maintenance by Service technician
Maintenance inter-
val
[1]
[1] D = day(s), W = week(s), M = month(s)
Maintenance work Information Note
1D 1W 1M 6M
()
b
Make a visual inspection
p. 53, §8.2
()
b
Perform adjustment
[2]
[2]
Operating Instructions for Control Unit
a
()( )()
b
Check/clean gas supply and discharge lines
[3]
[3] If available
a b
b
Check operating hours of DEFOR Analyzer module
[4]
[4] The UV lamp in the DEFOR Analyzer module must be replaced about every 2 years (
§8.1.2). With
measuring component NO: The gas filter fitted for NO measurement must be replaced every 2 years
approximately.
Note Explanation
a Maintenance interval depends on the individual application
b Only when solids deposit in the gas lines – as required
b
In addition, observe all official and internal company regulations valid for
the individual application.
Maintenance
interval
[1]
[1] M = month(s), Y = year(s)
Maintenance work Note
6M 1Y 2Y 10Y
❑❑
b
Check/service fitted gas pump
[2]
[2] Only for devices with Gas module with gas pump
a
❑❑
b
Check function of flow sensor
[3]
[3] Only for devices with Gas module with flow sensor
a
❑❑
b
Check important operating functions (e.g. alarm message) a
()
b
Readjust adjustment unit
[4]
[4] Only for devices with DEFOR Analyzer module with adjustment unit (option)
b
Adjust H
2
O measurement
[5]
[5] Only for devices with H
2
O measurement
b
Replace OXOR-E module
[6]
[6] Only for devices with OXOR-E Analyzer module
a
b
Check gas paths for leak tightness
b
Replace UV lamp
[7]
[7] Only for devices with DEFOR Analyzer module
b
Replace gas filter for NO measurement
[8]
[8] Only for devices with DEFOR Analyzer module and measuring component NO
b
Replace battery in control unit c
Note Explanation
a Maintenance interval depends on the individual application
c Carry out at the manufacturer's factory or workshop with suitable equipment