Operating Instructions FLOWSIC100
Gas Velocity Monitor
Product Description
8012513/03-2008 © SICK MAIHAK GmbH · Germany · All rights reserved 37
If a bus configuration is used with two connected sensors, the maximum cable length total
is reduced by half.
2.3.5 Connection Box
Connection cable master and slave have to be connected to the connection box. The
connection box separates the communication between master FLSE100 and MCU and
master FLSE100 and slave FLSE100.
2.3.6 Purge-air Unit Option
The purge-air unit is used to supply the sender/receiver units of the types FLSE100-PM, PH
and PHS with clean purged air.
Fig. 2.23: Purge-air unit SLV 1
Depending on the internal duct pressure, you may need to use additional reducers
(optional purge-air reducer set) or a purge-air unit with a more powerful fan.
Internal duct pressure (mbar) Reducer Fan type
-100 ... -20 40/7
2BH1300
-20 ... -10 40/10
-10 ... +30 -
+30 ... +100 - 2BH1400
Air filter
Fan
Baseplate
Purge-air hose