26 DUSTHUNTER T · Operating Instructions · 8012428 V 2.0 · © SICK MAIHAK GmbH Germany
Product Description
Control keys
2 I/O module
To be fitted on the module carrier, optional as:
2x analog output 0/4 ... 22 mA to output further measured variables (max. load 500
Ω)
2x analog input 0/4 ... 22 mA to read-in values from external sensors (gas tempera-
ture, internal duct pressure, moisture, O
2
) to calculate the dust concentration in
standard state.
3 Interface module
Module to pass on measured values, system status and service information to higher
level control systems, optionally for Profibus DP V0 or Ethernet, to plug onto hat rails. A
corresponding cable serves to connect the module to the connection board.
Key Function
Meas
Toggle between text and graphic display
Display the contrast setting (after 2.5 s)
Arrows Select next/previous measured value page
Status Display alarm or fault message
Menu Display main menu and selection of submenus
One module carrier is required for each module (to attach hat rails). One
module carrier is connected to the processor board using a special
cable and another one is docked.
For the DUSTHUNTER T50 version, a maximum of one analog input and
1 analog output module can be plugged on.
For the DUSTHUNTER T100/T200 version, a maximum of 1 analog
input module can be plugged on.
Profibus DP-V0 to transfer via RS485 according to DIN 19245 Part 3 as
well as IEC 61158.