FLOWSIC100 Operating Instructions
Gas Velocity Monitor
Commissioning and Configuration
110 © SICK MAIHAK GmbH · Germany · All rights reserved 8012513/03-2008
4.3 Extended Commissioning
4.3.1 Application Selection
See Section 2.2.2. The FLOWSIC100 offers means to measure on two measuring paths at the same time,
calculate one common measured value from them and output the common measured
value. This requieres 2 sender / receiver units per measuring path or one measuring probe
each (installation see Chapter 3). The necessary configurations are usually made in the
factory. If this is not the case (e.g. if existing devices are upgraded), carry out the following
procedure:
‡ In the window ‘‘Project Tree”, select the Typ MCU, change to User Access Level
‘‘Operator” and switch the system into the Maintenance Mode (see Section 4.2).
‡ Go to the submenu ‘‘Configuration / Application Selection”.
Fig. 4.29: Submenu ‘‘Configuration / Application Selection”
‡ In the window ‘‘Application Selection” select ‘‘FLOWSIC100 - 2 Path” in the field
‘‘Selection”.
‡ Close the SOPAS ET software and temporarily disconnect the MCU from the power supply
(power reset).
‡ If additional configuration is necessary, reconnect the measuring system to the SOPAS
ET software.
Note: Per default all measuring paths have the same weighting in the calculation of the output
value (for changing the weighting see Service Manual).