Operating Instructions FLOWSIC100
Gas Velocity Monitor
Commissioning and Configuration
8012513/03-2008 © SICK MAIHAK GmbH · Germany · All rights reserved 111
4.3.2 Configuring Optional Analog Modules
Module analog output
in addition max. 8 AO
available
The basic settings (subpage ‘‘Configuration / I/O Configuration / Output Parameter
FLOWSIC100”) apply to all available analog outputs.
To configure the settings, carry out the following procedure:
‡ Select the type MCU in the ‘‘Project Tree” window, enter the password for user level 1 and
switch the measuring system to Maintenance Mode (see Section 4.2).
‡ Go to the subpage ‘‘Configuration / I/O Configuration / Output Parameter FLOWSIC100”
(see Fig. 4.31).
‡ Activate the check box ‘‘Use first analog output module”. New dialog boxes for ‘‘Analog
Output 2 Parameter” and ‘‘Analog Output 3 Parameter” are opened.
‡ Configure the optional analog outputs according to the prerequisites in Section 4.2.3..
Fig. 4.30: ‘‘Configuration / I/O Configuration / Output Parameter FLOWSIC100” subpage
‡ If additional analog outputs have to be configured, activate the check box ‘‘Use second
analog output module”. New dialog boxes for ‘‘Analog Output 4 Parameter” and ‘‘Analog
Output 5 Parameter” are opened.
‡ To configure more analog outputs, continue as described above.