Operating Instructions FLOWSIC100
Gas Velocity Monitor
Commissioning and Configuration
8012513/03-2008 © SICK MAIHAK GmbH · Germany · All rights reserved 97
4.2.3 Configuring the Analog Output
Switch to the MCU subpage"Configuration / I/O Configuration / Output Parameter
FLOWSIC100" (see Fig. 4.13) and set the values for error and maintenance current.
Fig. 4.13: "Configuration / I/O Configuration / Output Parameter FLOWSIC100" subpage (example
for settings)
Field Parameter Notes
Analog output / general
configuration
Differential current Value > 20 mA
mA value to be output in case of mal-
functions
Output fault current
Yes The error current is output
No The error current is not output
Maintenance
current
Value if possible
≠
Live Zero
mA value to be output during Maintenance Mode
Current at
maintenance
Last value
The last measured value is output during Maintenance
Mode
User defined value
A value to be defined is output during Maintenance
Mode
Measured value
output
The current measured value is output during
Maintenance Mode