4.2.2 Configuring the Check Cycle
Basic Settings
‡ Select the type MCU on the "Network Scan Assistant" page and move it into the "Project
Tree" window (if not yet carried out).
‡ Select the type MCU in the "Project Tree" window and enter the password for user level 1.
‡ Switch the MCU to Maintenance Mode (see Section 4.2 ).
The output of the check cycle can be configured on the "Adjustment/Function Check -
Automatic" subpage (see Fig. 4.12). The function check can also be started manually.
Fig. 4.12: "Adjustment/Function Check - Automatic" subpage (example for settings)
Field Parameter Remark
Function control of
output duration
Value in seconds Output duration of the check cycle
Output control values
at AO
Inactive
The last measured value is output for the
duration of the check cycle.
Active
Check cycle is output at the analog
output
Check cycle interval Time between two check cycles See Section 2.5
Function control start
Hour
Specification of a start time in hours,
minutes and seconds
Minute
Second