FLOWSIC100 Operating Instructions
Gas Velocity Monitor
Commissioning and Configuration
112 © SICK MAIHAK GmbH · Germany · All rights reserved 8012513/03-2008
4.3.3 Configuring optional Interface Modules
The following steps are required for selection and configuration of the optionally available
interface modules Profibus DP and Ethernet:
‡ Select the type MCU in the "Project Tree" window, enter the password for user level 1
and switch the measuring system to Maintenance Mode (see Section 4.2).
‡ Switch to the "Configuration / I/O Configuration / System Configuration" subpage (see
Fig. 4.31). The installed interface module is displayed in the dialog box ‘‘Serial Extension
Module”.
‡ Configure the desired module according to the prerequisites.
Fig. 4.31: "Configuration / I/O Configuration / System Configuration" subpage
Note Two of the eight measured values of every measuring unit can be selected for the cyclical
data transfer if the Profibus DP module is installed.