FLOWSIC100 Operating Instructions
Gas Velocity Monitor
Commissioning and Configuration
118 © SICK MAIHAK GmbH · Germany · All rights reserved 8012513/03-2008
4.4.3 Configuration
Parameters for input / output (analog input / output) or installation (path length, installation
angle, duct diameter) can be changed with the following procedure:
‡ Go to the appropriate submen, select the line ‘‘Limit Low” or ‘‘Limit High” and press
‘‘Enter”
The valid range is diplayed in the fields ”Min“ and ”Max“
‡ Enter the default password ”1234“ with the keys ”^“ (scrolls from 0 to 9) and/or ”
→“ (mo-
ves the cursor right).
‡ Select the desired value for ”Min“ and ”Max“ with the keys ”^“ and/or ”
→“ and afirm with
”Save“.
The selected value is saved to the device.
Fig. 4.37: Menu structure for configuration
4.4.4 Change Application Parameter
‡ In the menu ‘‘I/O (MCU)” go into the submenu ‘‘I/O Parameter”, select the line ‘‘MCU
Variant” and affirm with ‘‘Enter”.
‡ In the submenu ‘‘MCU Variant”, select the line ‘‘FL100 2 path” and affirm with ‘‘Enter”.
Fig. 4.38: Menu structure for selection of two-path measurement
1 Path Length
2 Installation Angle
3 Cross Section Area
Parameter
/sensor/1/para/
Back Enter
1 Limit Low
2 Limit High
3 Live Zero
AO 1 Parameter
/io/para/ao/1
Back Enter
1 Limit Low
2 Limit High
AI 1 Parameter
/io/para/ai/1/
Back Enter
Limit Low
/io/para/ai/1/
Password
1234
/io/para/ai/1/
/sensor/1/para/
/io/para/ao/1/
Limit Low
/io/para/ao/1/
Back Save
∧→
Password
1234
Limit Low
/sensor/1/para/
Password
1234
0.000 Pa
6.2 m
0000.00 m³/h
Min:
Min:
Min:
Max:
Max:
Max:
Back Save
∧→
Back Save
∧→
1 AO Parameter
2 AI Parameter
3 MCU Variant
1 FL100 1 path
2 FL100 2 path
3 Unlimited
I/O Parameter MCU Variant
/io/para/ /io/para/var/
Back Enter
Back Enter