④ 1st STACK FACTOR PRESET’ s detail view
A. 1st STACK NUMB [1-99]: Stack number is to being transmitted.
B. ST HOST TRANS ENABLE: Yes/No for communication set-up of Stack to
arranged transmission target as host port.
C. STAND O2 DENSITY: Standard oxygen density of relevant Stack.
D. O2 COMPENS. LO-LIMIT: Calibration beginning section (function delete) that is
applied at oxygen calibration.
E. O2 COMPENS. HI-LIMIT: Calibration termination section (function delete) that is
applied at oxygen calibration.
F. ST PLANE AREA<A>(m3): Sectional area (m
2
) of relevant Stack at the point of
measuring instrument installation.
G. GAS DENSITY (kg/m3): Standard density of flue gas from relevant Stack
(kg/m
3
)
H. PITOT-TUBE FACTOR<C>: PITOT-Tube’ s coefficient of a flow meter.
I. ST HUMIDITY-<Xw> (%): Ratio of moisture that is included to flue gas from
relevant Stack (%)
[Picture 11]
⑤ 2nd STACK FASCTOR PRESET’ s detail view
The same as the 1st STACK FASCTOR PRESET’ s detail view.