5 MAINTENANCE
Lay the toothed belt on the drive axis, push pivoted shutter back onto the axis and
reinstall the tension spring.
Trigger the function check (select the “Adjustment / Function Check - Manual”
subdirectory in the “MCU“ device file and click “Start Manual Function Check”; see
“DUSTHUNTER T Parameter protocol (example)”, page 82).
In SOPAS ET, connect with “DH T200”, select the “Diagnosis / Check values” directory
and check the contamination value.
Fig. 71: SOPAS ET menu: DH T200/Diagnosis/Check values
If the measured values for contamination, zero point and span are within the allowed
ranges, save them to the device by clicking the “Refresh” button (“Check values” field); if
not, repeat cleaning and check the contamination value again by triggering a renewed
function check.
Put the pivoted shutter cover back on, remove the cover from the assembly flange,
swivel the enclosure back in and lock with the knurled screws.
Move the pivoted shutter back into the measuring position. To do this, click the
“Measurement” button under “Pivoted shutter reflector” in the “Adjustment / Manual
adjustment / Motor control” directory (see “SOPAS ET menu: DH T200/Adjustment/
Manual Adjustment/Motor control”, page 99).
Resume Measuring mode.
The contamination value can also be displayed on the LD display of the MCU
(trigger a function check and select the “T200/Diagnosis” menu, see “Menu
structure”, page 87).
If the contamination value does not sink below the value for warning in spite of
several cleaning processes, the device is probably defective contact SICK
Service.