99
8012428/YWL2/3-0/2016-08| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER T
Subject to change without notice
MAINTENANCE 5
DUSTHUNTER T200
▸ Set the sender/receiver unit to “Maintenance” mode (see “Cleaning the optical inter-
faces on the reflector”, page 99) and enter the Level 1 password.
▸ Loosen the knurled screws and swivel the enclosure to the side.
▸ Close the mounting flange with cover (see “Miscellaneous”, page 125).
▸ Click the “Mounting” button in menu “DH T200/Adjustment/Manual Adjustment/Motor
control” under position Pivoted shutter reflector” (see “SOPAS ET menu: DH T200/
Adjustment/Manual Adjustment/Motor control”, page 99).
The pivoted shutter then moves to the cleaning position.
Fig. 69: SOPAS ET menu: DH T200/Adjustment/Manual Adjustment/Motor control
▸ Remove pivoted shutter cover (1), loosen tension spring (6) and take pivoted shutter (2)
off axis (5).
▸ Carefully clean glass pane (4) (both sides) and reflector optics (3) with an optics cloth.
Fig. 70: Cleaning the optical interfaces on the reflector
Pivoted shutter cover
Pivoted shutter
Reflector optics
Glass pane (both sides)
Axis
Tension spring