99
8012428/YWL2/3-0/2016-08| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER T
Subject to change without notice
MAINTENANCE 5
DUSTHUNTER T200
Set the sender/receiver unit to “Maintenance” mode (see “Cleaning the optical inter-
faces on the reflector”, page 99) and enter the Level 1 password.
Loosen the knurled screws and swivel the enclosure to the side.
Close the mounting flange with cover (see “Miscellaneous”, page 125).
Click the “Mounting” button in menu “DH T200/Adjustment/Manual Adjustment/Motor
control” under position Pivoted shutter reflector” (see “SOPAS ET menu: DH T200/
Adjustment/Manual Adjustment/Motor control”, page 99).
The pivoted shutter then moves to the cleaning position.
Fig. 69: SOPAS ET menu: DH T200/Adjustment/Manual Adjustment/Motor control
Remove pivoted shutter cover (1), loosen tension spring (6) and take pivoted shutter (2)
off axis (5).
Carefully clean glass pane (4) (both sides) and reflector optics (3) with an optics cloth.
Fig. 70: Cleaning the optical interfaces on the reflector
Pivoted shutter cover
Pivoted shutter
Reflector optics
Glass pane (both sides)
Axis
Tension spring