65
8012428/YWL2/3-0/2016-08| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER T
Subject to change without notice
START-UP AND PARAMETER SETTINGS 4
Fig. 40: Cover screw for focussing opening (shown for DUSTHUNTER T100/T200)
▸ Insert the screw driver in the focussing opening and set the adjustment screw so that
the scale pointer in the control window points to the distance between the optical
surfaces of the sender/receiver unit and the reflector.
DUSTHUNTER T50:
Distance = dimension A + 250 mm (Fig. “Assembly on a dust-free path with adjusting
stands (shown for DUSTHUNTER T100)”)
DUSTHUNTER T100/T200:
Distance = dimension A + 326 mm (Fig. “Assembly on a dust-free path with adjusting
stands (shown for DUSTHUNTER T100)”)
Fig. 41: Focussing the sender light beam
▸ Swivel the electronics unit back to measuring position and lock with the knurled screws.
Pivoted shutter cover
Cover screw for focussing opening
The scale illumination lights when the measuring system is switched to “Mainte-
nance” or up to 10 min after device reboot.
4-quadrant display
Focussing opening
Scale
Control window
Reduce active measuring path
Electronics unit
Increase active measuring path