TROUBLESHOOTING AND CLEARING MALFUNCTIONS
9.3.1 Troubleshooting and clearing malfunctions, evaluation unit
Error indication Possible cause Clearance
LED Malfunction blinks;
(LED Warning could go on)
Relay 1:
Group malfunction
Plain-text error messages
indicate possible causes
Start Diagnosis mode (diag):
–Call up menu Error (or Warning)
Check and clear the specified malfunction.
Error description/message Component/possible cause Clearance
Evaluation unit not
responding
Evaluation unit:
Evaluation unit power supply
defective
Check power supply on all system components:
If necessary, provide power supply on site
If necessary, check/reconnect connections on
the system components
Evaluation unit:
Incorrect operating voltage
Check operating voltage set on the evaluation
unit:
If necessary, change setting
Evaluation unit:
Defective fuse
Check fuse in the evaluation unit:
If necessary, replace fuse
Evaluation unit:
No defect localized yet
Disconnect all system components from the
power supply and reconnect one at a time
Check the CAN bus cable from the evaluation
unit to the SR-unit resp. terminal box
Evaluation unit:
Error occurs again
Replace the last component connected, contact
Service
Evaluation unit:
24V/5V supply defective
Check 24V/5V supply, replace evaluation unit
resp. electronic board module; contact Service
Evaluation unit:
Inconsistent data detected in
parameter memory
Press Enter to restart the system; the factory
parameter settings are then active;
If necessary, reconfigure the parameters
If the same error message occurs again, replace
the EvU and contact Service