113
8009441/YN36/V3-0/2015-08| SICK OPERATING INSTRUCTIONS|GM35
Subject to change without notice
TECHNICAL DATA, CONSUMABLES AND SPARE PARTS
10 Technical Data, Consumables and Spare Parts
All the technical data are provisional specifications for GM35 applications. The GM35
analyzer is calibrated for specific applications once all the technical details have been
clarified.
10.1 Measuring components and accuracy
The minimum measuring range end values for the measuring components available with
the different device version are specified for a 1 m measuring gap. The maximum
measuring range end value for 1 m measuring gap is for
CO: 25.000
mg/m
3
N
2
O: 5.000 mg/m
3
H
2
O and CO
2
: 100 % by vol.
All the data refer to devices calibrated at the factory.
Combining measuring components
Minimum and maximum measuring range end values only apply to individual components.
Extremely small or large end values for different components cannot be combined in all
cases.
Measured Data
Measuring path Adaptable flange – flange distance 0.7…7.5 m
Larger distances on request
Measuring ranges For 1 m actual measuring path
CO: 0…225 mg/m
3
N
2
O: 0 ... 120 mg/m
3
CO
2
: 0…22.5 % by vol.
H
2
O: 0…25 % by vol.
Accuracy Stability relative to measuring range end value
Zero point: ± 2%
Sensitivity: ± 2% (in maintenance interval)
Suitability test EN 2001/80/EC and /EN 2000/76/EC
[1]
[1] Cross-Duct and GMP version for CO,CO
2
and H
2
O