EasyManua.ls Logo

SICK GM35

SICK GM35
130 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
80
8009441/YN36/V3-0/2015-08| SICKOPERATING INSTRUCTIONS|GM35
Subject to change without notice
START-UP
Check the display:
Adjust the two screws on the device flange, see “Aligning the optical axis”, page 92, until
the values
dx and dy on the display are less than or equal to ±0.1
(dx…value for horizontal position or adjustment)
(dy…value for vertical position or adjustment).
Values A, B, C and D are values of device-internal parameters. They should be in the
range 450…650.
If the valid range for A to D cannot be reached even after adjusting the screws several
times and the values for dx and dy are 0.1, check the warning and error messages in
Diagnosis mode, (
see “Menu structure”, page 43), and take the appropriate
maintenance or servicing measures. See Section see “Maintenance”, page 88.
Switch to Measuring mode:
Press meas.
7.5.3 Performing zero adjust
After the GM35 SR-unit and evaluation unit have been connected to the power supply, a
warming up time of approx. 2.5 - 5 hours (depending on ambient conditions) is required
before the zero adjust.
Allow the warming up phase (approx. 2.5 hours) to elapse after the SR-unit has been
connected to the power supply.
After the warming up phase, align the optical axis as described in “Fine adjustment on
the optical axis on the SR-unit”, page 79.
The following menu item is available in menu cal:
Zero Adjust – zero adjust (measuring path free from sample gas)
Activate Calibration mode (button “cal”), select menu item Zero Adjust and trigger with
Enter.
Enter the ambient temperature as prompted (accuracy ±2 °C).
Confirm the prompt. Zero adjust runs and Please wait is displayed.
When zero adjust has completed, either a positive confirmation or an error message is
displayed.
Adjustment successful
Switch to Measuring mode or define further parameter settings.
Adjustment not possible
An error message indicates that a malfunction occurred during the adjustment procedure.
The flashing “Malfunction” LED indicates that one or more error messages are pending.
Switch to Diagnosis mode and rectify the problem,
see “Troubleshooting and Clearing
Malfunctions”, page 104
.
cal

Table of Contents

Other manuals for SICK GM35

Related product manuals