START-UP
7.6.4 Installing the weatherproof cover for the SR-unit
The weatherproof cover is used when the measuring system is operated outdoors.
Fig. 65: Fitting the weatherproof cover for the GM35 SR-unit
See page 84 1 Bond the two spacer pins supplied centered on the flange fixture of the SR-unit at a
distance of 170 mm to each other, i.e. so that each pin is 85 mm from the center. Use a
quick-drying epoxy resin adhesive.
2 Seal the two unused mounting holes in the center of the weatherproof cover with
dummy screws.
3 Position the weatherproof cover on the SR-unit. The threads of the spacer pins now
protrude through the front fixing holes of the weatherproof cover.
4 Secure the weatherproof cover in position with the two knurled nuts.
7.6.5 Installing the weatherproof cover for the purge air units
The weatherproof cover for the purge air unit is available as an accessory. It comprises a
cover and a lock set.
GM35SR-unit
2 spacer pins,
secured with adhesive
Weatherproof
cover
2 knurled nuts
Washer
Spring washer
Screw
Nut
Spacer pins with knurled nuts