8009441/YN36/V3-0/2015-08| SICK OPERATING INSTRUCTIONS|GM35
Subject to change without notice
START-UP
Fig. 63: Cable connections on the purge air fixtures
Fig. 64: Cable connections on the SR-unit (underside)
Switch on the power supply for the SR-unit and the evaluation unit.
Carry out fine alignment according to
“Fine adjustment on the optical axis on the SR-
unit”, page 79.
Check once again that optical alignment is correct, preferably after having waited
approx. 30 minutes for the system to reach operating temperature.
To do this, call up menu Adjust Probe on the EvU and check the values displayed. See
page 79.
Reactivate Measuring mode:
Press meas
Regular measuring operation now starts.
CAN cable for SR-unit,
underside
CAN cable, SR-unit –
reflector
Low-pressure monitor
SLV of SR-unit
Low-pressure monitor
SLV of reflector
Temperature sensor
PT 1000
Power supply cable
CAN cable to purge
air fixture
CAN cable to EvU
Functional earth cable
(2.5 mm
2
)