20
8009441/YN36/V3-0/2015-08| SICKOPERATING INSTRUCTIONS|GM35
Subject to change without notice
PROJECT PLANNING INFORMATION
3.3 Initial onsite installation
The following work can be carried out by the customer's technicians. Requirement: The
project planning checklist has been processed beforehand.
3.3.1 Assembly preparation at the measuring point
This Section describes the welding work on the duct including making fixing elements
onsite.
Fig. 6: Fitting recommendation for the mounting flange (duct diameter not representative)
3.3.2 Uncovering the duct
If necessary, remove approx. 800 x 1500 mm (W x H) of the duct insulation to be able to
access the duct during subsequent work.
Keep the insulation material removed for later refitting resp. provide new suitable
insulation material.

Duct wall (steel)
Duct insulation
Flange with tube,
standard: L = 240 mm
Working platform
1.3 – 1.5 m
Distance to SLV4
min. 700 mm
Markings for fitting position, pointing
in flow direction
Active measuring path
Flange – flange distance:
CAUTION: Protective measures at the measuring point
Always shut down the installation before any work on the duct!
Secure parts to be separated with, for example, wire binding, to prevent
damage by falling objects.
Take appropriate protective measures against hot, explosive gases or toxic
gases that could possibly escape from the duct.
If necessary, seal off the mounting flange with a cover securely until device
assembly (e.g. for overpressure in duct).