CONNECTING THE SYSTEM CONTROL UNIT - SCU
Devices have been found; the connected GM35 is shown in the device list but is marked
with a red warning symbol “not available”. The device description file has not been
installed.
see “Load the device parameters:”, page 50.
Start a new project and then perform step “Perform a network scan:“.
If this is not successful, check the network configuration and set the following
standard values.
To do this, click the “Network Configuration” button and then the “Advanced” button
in the window that appears, and then enter the parameters as shown.
Perform a new network scan and follow the instructions from step “Perform a network
scan:“
.
6.1.4 Direct serial connection to the GM35 evaluation unit
Connecting the PC with SOPAS ET directly to the GM35 evaluation unit via a serial
interface cable:
Connect the interface cable to the Sub-D socket of the EvU (
“Connections on the
evaluation unit (cabling provided by the customer)”, page 38
) and to the Sub-D plug
of the PC.
Use the connection options according to
“Connecting the GM35 evaluation unit via the
SCU operating unit”, page 49
onwards:
1. Start SOPAS
2. Connect to the device
Confirm the selection “Connect to new device” in the “Welcome to SOPAS” with OK.
Follow the instructions of the Connection Wizard.
Use a serial interface cable (RS232) with a plug (Sub-D, 9-pole) and a socket
to connect the GM35 evaluation unit to a PC/laptop.