Fig. 13: Cable routing diagram
Power supply
(230/115 V AC) 4 m,
3 x 1.5 mm
2
Cable (CAN); 5, 10, 12, 15, 24 m
Evaluation unit
(cast enclosure shown)
CAN bus extension, ready for
connection, 15 m (option)
CAN bus cable
(standard), 4 m
Terminal box (option)
Power supply (230/115 V AC)
3 x 0.75 mm
2
3 binary inputs
6 x 0.5 mm
2
3 analog inputs
6 x 0.5 mm
2
3 binary outputs 6 x 0.5 mm
2
3 analog outputs 6 x 0.5 mm
2
Purge air unit SLV 4
(2 BH 13)
Power supply
4 x 1.5 mm
2
Filter monitor; 5 m
(2 and 3 m extensions possible)
to lengthen the CAN bus connection with
a cable
(1 x 2x 0.5mm
2
, twisted pair, shielded)
provided by the customer
Purge air unit SLV 4
(2 BH 14)
Temperature sensor
PT 1000
Cable connection, standard, included in
scope of delivery
Cable connection, option
Prefabricated with plug-in connector(s)
Onsite wiring
SR-unit with purge air fixture
Reflector with purge air fixture