START-UP
7.6.7 Setting up the SCU System Control Unit
The following index values serve to access the GM35 measured values and parameters.
Mapping Table
Measured values on SCU – Measured value (MV)
Control values on SCU – Monitor values (MO)
Operating state of the GM35 – State (S)
Diagnosis Table – DiagFlags (F01-F32, M01-M32)
Basic procedure and further information on the SCU
→ Operating Instructions “SCU”.
Index Measured value
MV01 H2O (g/m3, % by vol.
[1]
MV02 CO
or N
2
O
(mg/m3, ppm)
1)
MV03 CO2 (mg/m3, ppm)
1)
MV04 T (K)
MV05 P (hPa)
[1] Unit set during parametrization
Index Measured value
MO01 H2O_Zero (g/m3, % by vol.)
[1]
MO02 H2O_Span (%)
[2]
MO03 CO_Zero
or N
2
O
(mg/m3, ppm)
1)
MO04 CO_Span
or N
2
O
(%)
2)
MO05 CO2_Zero (g/m3, % by vol.)
1)
MO06 CO2_Span (%)
2)
[1] Unit set during parametrization
[2] Percentage deviation.
Index Operating state
S02 Measuring
S04 Maintenance
S04 RCycle
S05 CCycle
Index Diagnosis message (F ... Failure, M ... Maintenance)
F01 Sensor communication
F02 Incompatible device
F03 H2O EEPROM
F04 CO
(N
2
O)
EEPROM
F05 CO
(N
2
O)
com.
F06 Zero com.
F07 CO
(N
2
O)
CUVETTE com.
F08 H2O CUVETTE com.
F09 Filter com.
F010 VISOR com.
F011 CO
(N
2
O)
temp. detec.
F012 H2O temp. detec.
F013 CO
(N
2
O)
div. Zero
F014 H2O div. Zero
F015 VISOR fault