TROUBLESHOOTING AND CLEARING MALFUNCTIONS
H2O Reference Deviation of control value
measurement too high.
Carry out zero adjust on GM35;
see page 79.
Perform maintenance, see “Maintenance”,
page 88.
Contact Service if necessary.
HYGRO com No connection between CO
2
/H
2
O
module and humidity sensor.
Check cable connection is secure with correct
seat; repair if necessary.
Switch to default value.
If the error cannot be cleared: Contact Service.
HYGRO internal No communication with internal
humidity sensor.
Sensor defective/cable broken.
Check cable connections.
Exchange sensor.
MIRROR adj. Automatic beam tracking: Deviation of
optical axis is greater than area that
can be tracked.
Realign optical axis between SR-unit and
reflector,
page 92.
MR Adjustment Visor unit shows deviation (> 0.5 of
setpoint position).
Realign optical axis between SR-unit and
reflector; page 92.
Software version EvU and SR-unit incompatible. Install compatible (current) software for EvU
and SR-unit.
Contact Service if necessary.
VISOR amplifier Amplifier in visor at limit.
Amplifier regulation made in check
cycle; max. value 255.
Check contamination; clean optical interfaces
(
see page 90f).
Exchange IR source or contact Service
Call up menu DIAG/Sensor Values/VIS AMP
MS and check value,see “Checking the IR
source”, page 93
.