TECHNICAL DATA, CONSUMABLES AND SPARE PARTS
Purge air unit
Power supply
(three-phase)
Δ 200–240 V, Y 345–415 V at 50 Hz
Δ 220–275 V, Y 380–480 V at 60 Hz
Rated current Δ 2.6A / Y 1.5A at 50Hz
Δ 2.3A / Y 1.3A at 60Hz
Motor rating 0.35 kW at 50 Hz
0.45 kW at 60 Hz
Flow rate Min. 40 m
3
/h
Dimensions (WxHxD) 550 x 550 x 270 mm
Weight 14 kg
GM35 Evaluation unit (EvU)
Connections/interfaces
Data transmission
within the GM35
measuring system
CAN bus
• Max. line length 1000 m
• Electrically isolated
• Connects the EvU to the SR-unit
Service interface for PC RS 232
• Connection via 9-pole Sub-D socket
Analog outputs 3 pcs Output range: 0–20 mA, max. 500 Ω, electrically isolated, live zero
adjustable to 4 mA
A1–
A3
Measured values; assignment can be set individually
Analog input 0 ... 20 mA, 100 Ω
Status outputs 3 pcs Relays
DC max. 30 W, 48 V, 1 A
AC max. 60 VA, 48 V, 1 A
R1 Failure (NC contact)
R2 Maintenance request (NO contact)
R3 Function control (NO contact)
Status inputs 3 pcs Inputs for connecting potential-free contacts (loadable with 24 V;
supplied by GM35 evaluation unit)
E1 Check cycle
E2 Autocal in preparation
Power supply
Voltage/frequency 115/230 V AC ±10%; 50/60 Hz
Power input 50 VA max.
Dimensions, weight, protection class
Protection class IP 65 / NEMA 4x
Dimensions see dimension drawing from
page 118