Clinitek Advantus Operator’s Guide 61
Troubleshooting
The data includes
characters that
cannot be
transferred. The
characters that
can be
transferred are
those within
ASCII code range
0032 to 0126,
with the
exception of
these characters:
& \ ^ |
Ensure that the loadlist uses only
characters that can be transferred.
A run is in
progress or the
analyzer is not
displaying the
Ready/Run
screen when the
loadlist is
downloaded
Allow all tests in the current run to
complete and the analyzer to return
to the Ready/Run screen.
A loadlist has
already been
downloaded and
not all tests have
been run
Complete all tests on the current
loadlist before transferring another
loadlist.
When the problem that caused the
loadlist to fail is removed, send the
loadlist to the analyzer.
Error 01
Error 02
Error 03
Error 04
Error 05
Analyzer optical
error
1. Turn analyzer power off.
2. Wait several seconds.
3. Turn analyzer power on.
Symptom Possible Cause Remedy

Table of Contents

Other manuals for Siemens Clinitek Advantus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens Clinitek Advantus and is the answer not in the manual?

Siemens Clinitek Advantus Specifications

General IconGeneral
Test TypesUrinalysis
InterfaceRS-232, USB
CalibrationAutomatic calibration
Measurement PrincipleReflectance photometry
Sample TypeUrine
Power Supply100-240 VAC, 50/60 Hz
TypeUrine analyzer
DisplayTouchscreen
Reagent TypeTest strips
Measurement ParametersGlucose, Bilirubin, Ketone, Specific Gravity, Blood, pH, Protein, Urobilinogen, Nitrite, Leukocytes

Related product manuals