EasyManuals Logo

Siemens Clinitek Advantus User Manual

Siemens Clinitek Advantus
148 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #62 background imageLoading...
Page #62 background image
62 Clinitek Advantus Operator’s Guide
Troubleshooting
Error 06-2 A reagent strip
detected at the
first readhead is
not detected at
the second
readhead
1. Select
Return to Ready/Run to
cancel the run and return to the
Ready/Run screen.
2. Turn analyzer power off.
3. Remove the fixed platform to
locate the strip.
4. Check the pins on the moving
table to ensure that none are
bent or broken.
5. Perform the Performing the
Daily Cleaning‚ page 37.
6. Check your printout of results,
or the Results Error Report to
determine the specimen(s) for
which no results exist.
7. Retest those specimens.
Error 07-1 A reagent strip
either is not fully
wetted or is
upside-down on
the platform
1. If the strip is upside-down,
remove and clean the push bar,
the fixed platform, and the
holddown plate.
2. Check your printout of results,
or the Results Error Report, to
determine the specimen(s) for
which no results exist.
3. Retest those specimens.
Ensure that the strip is dipped
completely into the specimen
and is placed onto the platform
with the pads facing up.
Symptom Possible Cause Remedy

Table of Contents

Other manuals for Siemens Clinitek Advantus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens Clinitek Advantus and is the answer not in the manual?

Siemens Clinitek Advantus Specifications

General IconGeneral
Test TypesUrinalysis
InterfaceRS-232, USB
CalibrationAutomatic calibration
Measurement PrincipleReflectance photometry
Sample TypeUrine
Power Supply100-240 VAC, 50/60 Hz
TypeUrine analyzer
DisplayTouchscreen
Reagent TypeTest strips
Measurement ParametersGlucose, Bilirubin, Ketone, Specific Gravity, Blood, pH, Protein, Urobilinogen, Nitrite, Leukocytes

Related product manuals