8-5
1010PFM-3G
Section 8
Press the key to select the Calibration menu.
Press the key to select the Automatic Calibration menu.
Now you have to place the 1011TS transducer flush onto the [0.100] calibration block on the transducer
holder:
a.
A
pply a small amount of coupling
compound to the surface of the
calibration block.
b.
Place the transducer firmly against
the built-in calibration block.
c.
Make sure that the transducer face is
centered on the calibration block.
Press the key to edit the [Set Xdcr Zero] value. Note that an equal (=) sign appears before the
number: [=0.100]
Press the key to move the highlight to Calibration.
Measure Thickness
Calibration
Automatic
Manual
Set Xdcr Zero 0.100
Select Material Steel
Calibrate Material 0.500
Set Xdcr Zero =0.100
Select Material Steel
Calibrate Material 0.500