8-6
1010PFM-3G
Section 8
Holding the transducer against the calibration block, press to calibrate the [Set Xdcr Zero] value.
The thickness screen shows the 0.100 measurement (similar to figure below):
Now you have to place the 1011TS transducer flush on to the [0.500] calibration block on the trans-
ducer holder. Press to set the [Calibrate Material] value.
Press the key to return to the menu. This completes the calibration procedure. The 1011TS thick-
ness gauge transducer is now certified for operation with the flow computer.
z Repeat the calibration procedure for best accuracy.
z The thickness screen shows the 0.500 measurement (similar to figure above). This cali-
brates the 1011TS satisfactorily for most pipe materials. If you have reason to suspect its
accuracy on your material, try the calibration procedure using a sample of your pipe material
of a known thickness.
z If your pipe material is not listed in the “select material” menu, then refer to paragraph 8.6
“Manual Calibration of Thickness Gauge.”
NOTES
ENT
ENT
Press the key to return to the calibration menu. Use the to move the highlight to [Calibrate
Material], then press the to enable editing. Note that an equal (=) sign appears before the num-
ber: [=0.500].
The thickness screen shows the 0.500 measurement (similar to figure below):
Siemens
Siemens
MENU
MENU
Set Xdcr Zero 0.100
Select Material Steel
Calibrate Material =0.500