4.2.2 Sensor installation
4.2.2.1 Determining a location
CAUTION
Electromagnetic fields
Do not install the flowmeter in the vicinity of strong electromagnetic fields, for example near
motors, variable frequency drives, transformers etc.
Upstream / downstream
No pipe run requirements, that is straight inlet/outlet sections, are necessary.
Avoid long drop lines downstream from the sensor to prevent process media separation
causing air / vapor bubbles in the tube (min. back pressure: 0.2 bar).
Avoid installing the flowmeter immediately upstream of a free discharge in a drop line.
Location in the system
The optimum location in the system depends on the application:
Liquid applications
Gas or vapor bubbles in the fluid may result in erroneous measurements, particularly in the
density measurement.
Do not install the flowmeter at the highest point in the system, where bubbles will be
trapped.
Install the flowmeter in low pipeline sections, at the bottom of a U-section in the pipeline.
Figure 4-2 Liquid applications, wrong location with trapped air/gas
Gas applications
Vapor condensation or oil traces in the gas may result in erroneous measurements.
Do not install the flowmeter at the lowest point of the system.
Install a filter.
Figure 4-3 Gas applications, wrong location with trapped oil
Installing/mounting
4.2 Installation instructions
SITRANS FC330
Operating Instructions, 01/2019, A5E44030648-AB 49