8.2.6 Inputs and outputs
The hardware functionality of input and output is fixed when ordering the product. The available
configuration is described in the following table:
Channel HW configuration
(fixed when ordering)
SW configuration
available to the user
1 Current output (4-20 mA)
HART, Modbus, PROFIBUS
DP, PROFIBUS PA
2 Signal output
Current (0/4-20 mA)
Frequency or pulse
Three-stage analog valve dosing control
Discrete one or two-valve dosing control
Operational and alarm status
3 Signal output
Current (0/4-20 mA)
Frequency or pulse
Redundant frequency or pulse (together with channel 2)
Three-stage analog valve dosing control
Discrete one or two-valve dosing control
Operational and alarm status
Relay output
Discrete one or two-valve dosing control
Operational and alarm status
Signal input
Dosing control
Totalizer reset
Remote zero adjust
Force or freeze output(s)
Parameter assignment