EasyManua.ls Logo

Siemens SITRANS FC330 - Connecting the Modbus (CH1)

Siemens SITRANS FC330
280 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Note
For Ex versions active or passive current output is preselected at ordering and cannot be
changed.
Non-Ex versions can be connected as either active or passive.
Note
Load
Signal output: < 500 Ω at 14 to 24 VDC (active), 14 to 30 VDC (passive)
Relay output: 30 VAC/VDC, 100 mA
Passive signal input: 15 to 30 VDC, 2 to 15 mA
5.2.2.4 Connecting the Modbus (CH1)
1. Remove cap and ferrule from cable gland and slide onto cable.
2. Push cable through open gland and cable path.
3. Restore ferrule and tighten cap to lightly hold cable in place.
Connecting
5.2 Connecting FC330
SITRANS FC330
Operating Instructions, 01/2019, A5E44030648-AB 69

Table of Contents

Related product manuals