Set date and time
4XLFNFRPPLVVLRQLQJ
About
Language
<HV
1R
&RQWLQXHZLWK4XLFNFRPPLVVLRQLQJ:L]DUG
)LUVW6WDUWXS:L]DUGLVQRZILQLVKHG
View
no.
Text Options/description
1 Language Set the language: English, Deutsch, Français, Italiano, Español,
2 About Information about the Quick commissioning wizard
3 Set date and time The set date and time (real time clock) is used for all time stamps of
logged information
4 Quick commissioning The Quick commissioning wizard comprises the most important pa‐
rameters/menus for quick configuration of the flowmeter
6.3.4 Zero point adjustment
The flowmeter system is optimized through a zero point adjustment which is performed via
the Zero point adjustment wizard (Page 90).
Performing a zero point adjustment
CAUTION
Gas application
Zero point adjusting the device is only recommended for liquid applications.
Commissioning
6.3 Local commissioning via HMI
SITRANS FC330
86 Operating Instructions, 01/2019, A5E44030648-AB