11.2 Cleaning
11.2.1 Cleaning the enclosure
Cleaning the enclosure
Clean the outside of the enclosure with the inscriptions and the display window using a
cloth moistened with water or a mild detergent.
Do not use any aggressive cleansing agents or solvents, e.g. acetone. Plastic parts or the
painted surface could be damaged. The inscriptions could become unreadable.
WARNING
Electrostatic charge
Risk of explosion in hazardous areas if electrostatic charges develop, for example, when
cleaning plastic surfaces with a dry cloth.
Prevent electrostatic charging in hazardous areas.
11.2.2 Servicing the remote seal measuring system
The remote seal measuring system usually does not need servicing.
If the mediums are contaminated, viscous or crystallized, it could be necessary to clean the
diaphragm from time to time. Use only a suitable solvent to remove the deposits from the
diaphragm. Do not use corrosive cleaning agents. Prevent the diaphragm from getting
damaged due to sharp-edged tools.
NOTICE
Improper cleaning of diaphragm
Device damage. The diaphragm can be damaged.
Do not use sharp or hard objects to clean the diaphragm.