5. Open the shut-off valve at the pressure tapping point.
6. Open the shut-off valve for the process .
8.5.1.2 Commissioning with steam or liquid
Requirement
All valves are closed.
Procedure

Pressure transmitter
Shut-off valve
Shut-off valve to process
Shut-off valve for test connection or for bleed screw
Pressure line
Shut-off valve
Drain valve
Compensation vessel (steam only)
To commission the pressure transmitter for steam or liquid, proceed as follows:
1. Open the shut-off valve for the test connection .
2. Via the test connection of the shut-off valve , apply the pressure corresponding to the
start of scale value to the pressure transmitter.
3. Ensure that the lower range value corresponds to the desired value. Otherwise, correct the
value.
4. Close the shut-off valve for the test connection .
Commissioning
8.5 Application examples
SITRANS P320/P420 (mA/HART)
90 Operating Instructions, 06/2018, A5E44852162-AA